    Quantitative comparison of PET performance—Siemens Biograph mCT and mMR

    • Original research
    • Open access
    • Published:

    EJNMMI Physicsvolume 3, Article number: 5 (2016) Cite this article

    Abstract

    Background

    Integrated clinical whole-body PET/MR systems were introduced in 2010.

    In order to bring this technology into clinical usage, it is of great importance to compare the performance with the well-established PET/CT. The aim of this study was to evaluate PET performance, with focus on image quality, on Siemens Biograph mMR (PET/MR) and Siemens Biograph mCT (PET/CT).

    Methods

    A direct quantitative comparison of the performance characteristics between the mMR and mCT system was performed according to National Electrical Manufacturers Association (NEMA) NU 2-2007 protocol.

    Spatial resolution, sensitivity, count rate and image quality were evaluated. The evaluation was supplemented with additional standardized uptake value (SUV) measurements.

    Results

    The spatial resolution